Published October 1, 2007 | Version v1
Journal article

Compton imager based on a single planar segmented HPGe detector

  • 1. Department of Physics, Royal Institute of Technology, AlbaNova University Center, Roslagstullsbacken 21, SE-106 91 Stockholm (Sweden)

Description

A collimator-free Compton imaging system has been developed based on a single high-purity germanium detector and used to generate images of radioactive sources emitting γ rays. The detector has a planar crystal with one pixellated contact with a total of 25 segments. Pulse shape analysis has been applied to achieve a 3D-position sensitivity of the detector. The first imaging results from this detector are presented, based on the reconstruction of events where a γ ray is fully absorbed after scattering between adjacent segments

Availability note (English)

Available from http://dx.doi.org/10.1016/j.nima.2007.06.065

Additional details

Identifiers

DOI
10.1016/j.nima.2007.06.065;
PII
S0168-9002(07)01326-5;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
580
Journal Issue
2
Journal Page Range
p. 1075-1078
ISSN
0168-9002
CODEN
NIMAER

Conference

Title
3. international conference on imaging techniques in subatomic physics, astrophysics, medicine, biology and industry
Dates
27-30 Sep 2006
Place
Stockholm (Sweden)

INIS

Optional Information

Copyright
Copyright (c) 2007 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.