Published October 2010 | Version v1
Journal article

Development of two-channel prototype ITER vacuum ultraviolet spectrometer with back-illuminated charge-coupled device and microchannel plate detectors

  • 1. National Fusion Research Institute, Gwahangno 113, Yuseong-gu, Daejeon 305-333 (Korea, Republic of)
  • 2. Institut fuer Plasmaphysik, Forschungszentrum Juelich Gmbh, D-52425 Juelich (Germany)
  • 3. ITER Organization, CS 90 046, 13067 St. Paul Lez Durance Cedex (France)

Description

A vacuum ultraviolet (VUV) spectrometer of a five-channel spectral system is designed for ITER main plasma impurity measurement. To develop and verify the system design, a two-channel prototype system is fabricated with No. 3 (14.4-31.8 nm) and No. 4 (29.0-60.0 nm) among the five channels. The optical system consists of a collimating mirror to collect the light from source to slit, two holographic diffraction gratings with toroidal geometry, and two different electronic detectors. For the test of the prototype system, a hollow cathode lamp is used as a light source. To find the appropriate detector for ITER VUV system, two kinds of detectors of the back-illuminated charge-coupled device and the microchannel plate electron multiplier are tested, and their performance has been investigated.

Additional details

Identifiers

Publishing Information

Journal Title
Review of Scientific Instruments
Journal Volume
81
Journal Issue
10
Journal Page Range
p. 10E508-10E508.3
ISSN
0034-6748
CODEN
RSINAK

Optional Information

Notes
(c) 2010 American Institute of Physics