Published September 1982 | Version v1
Journal article

Fluorescent x-ray analysis of iron content in iron ores by glass bead technique

  • 1. Sumitomo Metal Industries Ltd., Amagasaki, Hyogo (Japan). Central Research Labs.

Description

Fluorescent X-ray analysis of iron contents in iron ores by glass bead technique has been theoretically studied. Calculation has been made on FeK sub(α) fluorescent X-ray intensity in Fe-O binary sample and in Fe-O-impurity ternary sample in various fusion procedures by means of theoretical formulas for fluorescent X-ray intensity. The results are as follows: (1) In the comparison of characteristics of various fusion procedures, the intensity decreases, the derivative of calibration curve increases and matrix effect decreases with X-ray absorption of glass bead. For example, in the procedure in which lanthanum oxide is used as reagent, matrix effect is low and lanthanum oxide is useful as heavy absorber, while vanadium oxide has a little effect as heavy absorber and matrix effect is not low. (2) In the procedure in which anhydrous sodium tetraborate is used as flux, calibration curve for Fe-O binary system sample is made, and FeK FeK sub(α) intensity is calculated for Fe-O-impurity ternary system samples when impurity is substituted with oxygen in Fe-O binary system sample. Theoretical correction factors are derived from the change of FeK sub(α) intensity. It is shown in experiments that these correction factors have an effect on analysis of iron content. (author)

Additional details

Publishing Information

Journal Title
Trans. Iron Steel Inst. Jpn.
Journal Volume
22
Journal Issue
9
Series
Trans. Iron Steel Inst. Jpn.
Journal Page Range
724-731
ISSN
0021-1583