Published 1992
| Version v1
Journal article
Alpha scintillation cell for direct measurement of indoor radon
Description
A large volume (1500 cm3) alpha scintillation cell to measure indoor radon is described. Air is sampled directly into the cell and gross alpha activity is measured after three hours. The cells are suitable for concentrations higher than 10-20 Bq/m3. They were successfully used for randon measurements in kindergartens in Nova Gorica. 7 refs., 2 figs., 1 tab
Additional details
Publishing Information
- Journal Title
- Journal of Environmental Science and Health. Part A, Environmental Science and Engineering
- Journal Volume
- 27
- Journal Issue
- 6
- Journal Page Range
- p. 1535-1540.
- ISSN
- 0360-1226
- CODEN
- JESEDU
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 26020524
- Subject category
- S54: ENVIRONMENTAL SCIENCES;
- Descriptors DEI
- AIR POLLUTION MONITORS; ECOLOGICAL CONCENTRATION; INDOOR AIR POLLUTION; MONITORING; RADON; SCINTILLATION COUNTERS
- Descriptors DEC
- AIR POLLUTION; ELEMENTS; MEASURING INSTRUMENTS; MONITORS; NONMETALS; POLLUTION; RADIATION DETECTORS; RARE GASES