Published 2015 | Version v1
Miscellaneous Open

Characterization and standardized radiation qualities deployment - NBR IEC 61225 - in a animal irradiator

  • 1. Universidade do Estado do Rio de Janeiro (LCR/UERJ), Rio de Janeiro, RJ (Brazil). Lab. de Ciencias Radiologicas
  • 2. Instituto de Radioprotecao e Dosimetria (IRD/CNEN-RJ), Rio de Janeiro, RJ (Brazil)

Description

The present work consisted in the determination of additional aluminum filters, so that the Animal Irradiator of the Laboratory of Radiological Sciences-UERJ can operate in accordance with standard IEC 61267: 2005. In order to perform the measurements necessary for the determination of these filters, it was developed an additional filter positioning system coupled to the X-ray tube. The determination of these filters was performed to obtain the first half value layer (HVL) according to the window with the values given by the IEC 61267: 2005. It was used IPEN SRS-78 (Institute of Physics and Engineering in Medicine) software for the estimation of the total amount of the additional filtration through the measurements values of the first HVL. (author)

Files

47080155.pdf

Files (45.9 kB)

Name Size Download all
md5:d144959b10c17fc9f228b5a230d9cd89
45.9 kB Preview Download

Additional details

Additional titles

Original title (Portuguese)
Caracterizacao e implantacao de qualidades de radiacao pradronizadas - NBR IEC 61225 - em um irradiador de animais

Publishing Information

Imprint Pagination
5 p.
Report number
INIS-BR--16417

Conference

Title
2. Brazilian congress on metrology of ionizing radiations
Acronym
CMRI 2015
Dates
29 Nov - 4 Dec 2015
Place
Bento Goncalves, RS (Brazil)