A low background gamma ray spectrometer with anticosmic shielding
- 1. University of Natural Sciences, Vietnam National University, Ho Chi Minh City (Viet Nam)
Description
The article describes a gamma ray spectrometer protected by a lead shield (Model 747E Canberra lead shield) and an active shield made of an 80 cm x 80 cm x 3 cm plastic scintillator plate in anticoincidence on top of the lead shield. The detector used as low background gamma-spectrometer is a high purity Germanium crystal of model GC2018 Canberra. The background count rate currently achieved (30-2400 keV) is 1.27 cps without anticoincidence. The level of background suppression obtained from the active protection is 0.80 overall and about 0.43 for the 511 keV gamma line. The gamma ray spectrometer is installed and operated in the Nuclear Laboratory, Department of Nuclear Physics, University of Science, Ho Chi Minh City Vietnam National University. (author)
Additional details
Publishing Information
- Journal Title
- Communications in Physics
- Journal Volume
- 26
- Journal Issue
- 1
- Series
- Published by the Vietnamese Academy of Science and Technology
- Journal Page Range
- p. 93-97
- ISSN
- 0868-3166
INIS
- Country of Publication
- Viet Nam
- Country of Input or Organization
- Viet Nam
- INIS RN
- 48103507
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ANTICOINCIDENCE; DESIGN; DIAGRAMS; EDUCATIONAL FACILITIES; GAMMA SPECTROMETERS; HIGH-PURITY GE DETECTORS; PLASTIC SCINTILLATORS; SHIELDING
- Descriptors DEC
- GE SEMICONDUCTOR DETECTORS; INFORMATION; MEASURING INSTRUMENTS; PHOSPHORS; RADIATION DETECTORS; SEMICONDUCTOR DETECTORS; SPECTROMETERS
Optional Information
- Notes
- 11 refs, 2 figs, 1 tab