Published November 2008
| Version v1
Journal article
A convenient alignment approach for x-ray imaging experiments based on laser positioning devices
- 1. Center for Bioengineering and School of Electrical and Computer Engineering, University of Oklahoma, Norman, Oklahoma 73019 (United States)
- 2. Department of Radiology, University of Alabama at Birmingham, Birmingham, Alabama 35233 (United States)
Description
This study presents a two-laser alignment approach for facilitating the precise alignment of various imaging and measuring components with respect to the x-ray beam. The first laser constantly pointed to the output window of the source, in a direction parallel to the path along which the components are placed. The second laser beam, originating from the opposite direction, was calibrated to coincide with the first laser beam. Thus, a visible indicator of the direction of the incident x-ray beam was established, and the various components could then be aligned conveniently and accurately with its help.
Additional details
Identifiers
- DOI
- 10.1118/1.2989059;
Publishing Information
- Journal Title
- Medical Physics
- Journal Volume
- 35
- Journal Issue
- 11
- Journal Page Range
- p. 4907-4910
- ISSN
- 0094-2405
- CODEN
- MPHYA6
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 44053385
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ALIGNMENT; BEAM POSITION; EQUIPMENT; IMAGES; LASERS; POSITIONING; RADIOTHERAPY; X RADIATION
- Descriptors DEC
- ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; MEDICINE; NUCLEAR MEDICINE; RADIATIONS; RADIOLOGY; THERAPY
Optional Information
- Notes
- (c) 2008 American Association of Physicists in Medicine