Problems and progress in the preparation of sources for the alpha spectrometry of plutonium

  • 1. IAEA, Vienna, Austria

Description

The interpretation of non-destructive measurements of plutonium materials require more accurate determinations of the isotopic abundance of Pu-238 than conventional chemical assays. The requirements of calorimetry, passive neutron and conventional chemical assays are presented and compared. When Pu-238 is measured by alpha spectrometry, these requirements define how well the plutonium must be separated from americium, and what should be the accuracy of the spectrometry. The latter can strongly depend upon the resolution of the alpha spectrum. The authors describe a procedure to produce sources by drop deposition which ensure a resolution of 17 keV with commercial instrumentation

Files

14729514.pdf

Files (507.3 kB)

Name Size Download all
md5:7d9feb53a52efc71debb2c64de42af64
507.3 kB Preview Download

Additional details

Publishing Information

Imprint Title
Minutes of the fourth SALE program participants meeting
Journal Page Range
p. 321-350.
Report number
CONF-810738--

Conference

Title
4. SALE program participants meeting.
Dates
Jul 1981.
Place
Argonne, IL (USA).