Published February 1986 | Version v1
Journal article

A method for calibration of cylindrical tissue equivalent proportional counters

  • 1. GE Reuter-Stokes, Edison Park, 8499 Darrow Road, Twinsburg, OH 44087

Description

A computer simulation technique was used to generate estimates of the pulse height distribution arising from (n,p) reactions in a cylindrical tissue equivalent radiator. The model assumes a mono-directional neutron beam interacts at a point on the cylinder and the resulting protons completely traverse the tissue equivalent gas volume. Features in the simulation model spectrum are correlated to spectral features produced by an experimental TEPC irradiated by an isotopic neutron source. Model predictions are also compared to actual spectra produced by using proton beams from a cyclotron. These three estimates for the energy calibration of the cylindrical TEPC are in good agreement. A simple method for energy calibration of cyindrical TEPC's will facilitate measurement of absorbed dose and dose equivalent due to neutrons. The use of TEPC based instruments would permit the refinement of personnel neutron dosimetry programs as performed for the nuclear power industry, medical neutron therapy, and research laboratories using radio-active sources

Additional details

Publishing Information

Journal Title
IEEE Trans. Nucl. Sci.
Journal Volume
NS-33
Journal Issue
1
Series
IEEE Trans. Nucl. Sci.
Journal Page Range
vp.
ISSN
0018-9499
CODEN
IETNA

Conference

Title
IEEE nuclear science symposium.
Dates
23-25 Oct 1985.
Place
San Francisco, CA (USA).

Optional Information

Secondary number(s)
CONF-851009--.