Published December 2002
| Version v1
Journal article
A new configuration of the Moxon-Rae detector based on Si detector
Creators
Description
A new Moxon-Rae detector configuration based on Si semiconductor detector was proposed in this paper. Three γ-ray sources, 137Cs, 60Co, and 24Na, were employed to make actual measurements using the new Moxon-Rae detector. The measured pulse height spectra and detection efficiencies were compared with the EGS4 simulated values. The results revealed that the proposed new configuration is indeed a successful method and specially a useful technique for higher energy γ-ray measurement
Additional details
Identifiers
- PII
- S1350448702001075;
Publishing Information
- Journal Title
- Radiation Measurements
- Journal Volume
- 35
- Journal Issue
- 6
- Journal Page Range
- p. 557-560
- ISSN
- 1350-4487
- CODEN
- RMEAEP
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 34008073
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- CESIUM 137; CONFIGURATION; EFFICIENCY; PERFORMANCE TESTING; SI SEMICONDUCTOR DETECTORS; SODIUM 24
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; CESIUM ISOTOPES; HOURS LIVING RADIOISOTOPES; INTERMEDIATE MASS NUCLEI; ISOMERIC TRANSITION ISOTOPES; ISOTOPES; LIGHT NUCLEI; MEASURING INSTRUMENTS; MILLISEC LIVING RADIOISOTOPES; NUCLEI; ODD-EVEN NUCLEI; ODD-ODD NUCLEI; RADIATION DETECTORS; RADIOISOTOPES; SEMICONDUCTOR DETECTORS; SODIUM ISOTOPES; TESTING; YEARS LIVING RADIOISOTOPES
Optional Information
- Copyright