Measurement for the MLC leaf velocity profile by considering the leaf leakage using a radiographic film
Creators
- 1. Department of Radiation Oncology, University of Toronto and Radiation Medicine Program, Princess Margaret Hospital, University Health Network, 610 University Avenue, Toronto, ON M5G 2M9 (Canada)
- 2. Medical Physics Department, Grand River Regional Cancer Center, Grand River Hospital, PO Box 9056, 835 King Street West, Kitchener, ON N2G 1G3 (Canada)
Description
A method to measure the velocity profile of a multi-leaf collimator (MLC) leaf along its travel range using a radiographic film is reported by considering the intra-leaf leakage. A specific dynamic MLC field with leaves travelling from the field edge to the isocentre line was designed. The field was used to expose a radiographic film, which was then scanned, and the dose profile along the horizontal leaf axis was measured. The velocity at a sampling point on the film can be calculated by considering the horizontal distance between the sampling point and the isocentre line, dose at the sampling point, dose rate of the linear accelerator, the total leaf travel time from the field edge to isocentre line and the pre-measured dose rate of leaf leakage. With the leaf velocities and velocity profiles for all MLC leaves measured routinely, a comprehensive and simple QA for the MLC can be set up to test the consistency of the leaf velocity performance which is essential to the IMRT delivery using a sliding window technique. (note)
Availability note (English)
Available online at http://stacks.iop.org/0031-9155/51/N299/pmb6_17_n02.pdf or at the Web site for the journal Physics in Medicine and Biology (ISSN 1361-6560) http://www.iop.org/Additional details
Identifiers
- URL
- http://stacks.iop.org/0031-9155/51/N299/pmb6_17_n02.pdf; http://www.iop.org/;
- DOI
- 10.1088/0031-9155/51/17/N02;
- PII
- S0031-9155(06)21385-X;
Publishing Information
- Journal Title
- Physics in Medicine and Biology
- Journal Volume
- 51
- Journal Issue
- 17
- Journal Page Range
- p. N299-N306
- ISSN
- 0031-9155
- CODEN
- PHMBA7
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 38008845
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BIOMEDICAL RADIOGRAPHY; COLLIMATORS; DOSE RATES; LINEAR ACCELERATORS; PERFORMANCE; RADIATION DOSES
- Descriptors DEC
- ACCELERATORS; DIAGNOSTIC TECHNIQUES; DOSES; MEDICINE; NUCLEAR MEDICINE; RADIOLOGY