Published February 1987 | Version v1
Journal article

Analysis of dose distributions for 4 MV X rays for symmetrical and offset fields using tapered multi-leaf collimators

  • 1. Japanese Foundation for Cancer Research, Tokyo. Cancer Inst.

Description

Dose distributions for 4 MV X rays for symmetrical and offset fields using tapered multi-leaf collimators were measured with a computer controlled dosimeter system and analysis of the obtained data was carried out. Off-center axis ratios (OCR) for square and rectangular fields were different in the region of low dose. The equation for the correction was shown. OCR's for offset fields were almost the same as those for symmetrical fields when the field widths were less than about 10 cm and the depths were greater than about 6 cm. (author)

Additional details

Publishing Information

Journal Title
Hoshasen Chiryo Shisutemu Kenkyu
Journal Issue
suppl.4
Series
Hoshasen Chiryo Shisutemu Kenkyu.
CODEN
HCSKE