Published February 1987
| Version v1
Journal article
Analysis of dose distributions for 4 MV X rays for symmetrical and offset fields using tapered multi-leaf collimators
Creators
- 1. Japanese Foundation for Cancer Research, Tokyo. Cancer Inst.
Description
Dose distributions for 4 MV X rays for symmetrical and offset fields using tapered multi-leaf collimators were measured with a computer controlled dosimeter system and analysis of the obtained data was carried out. Off-center axis ratios (OCR) for square and rectangular fields were different in the region of low dose. The equation for the correction was shown. OCR's for offset fields were almost the same as those for symmetrical fields when the field widths were less than about 10 cm and the depths were greater than about 6 cm. (author)
Additional details
Publishing Information
- Journal Title
- Hoshasen Chiryo Shisutemu Kenkyu
- Journal Issue
- suppl.4
- Series
- Hoshasen Chiryo Shisutemu Kenkyu.
- CODEN
- HCSKE
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 18082714
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- COLLIMATORS; COMPUTER CALCULATIONS; DEPTH DOSE DISTRIBUTIONS; EXTERNAL IRRADIATION; IRRADIATION PROCEDURES; PHANTOMS; RADIOTHERAPY; X RADIATION; X-RAY DOSIMETRY
- Descriptors DEC
- DISTRIBUTION; DOSIMETRY; ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; IRRADIATION; MEDICINE; MOCKUP; RADIATION DOSE DISTRIBUTIONS; RADIATIONS; SPATIAL DISTRIBUTION; SPATIAL DOSE DISTRIBUTIONS; STRUCTURAL MODELS; THERAPY