Development of an orthogonal transient internal probe

Description

The transient internal probe (TIP) diagnostic is designed to measure the magnetic field profile internal to hot, high density plasmas. A small magneto-optic probe is fired through the plasma at velocities in excess of 2 km/s. Magnetic fields are obtained by illuminating the probe from the front with an argon ion laser and measuring the amount of Faraday rotation in the reflected beam. First generation TIP probes are limited in that only one component of the magnetic field can be measured. The orthogonal TIP probe, however, will have the capability of measuring magnetic fields which are perpendicular to the direction of probe travel. Orthogonal measurement capability can provide a full radial scan of the poloidal field profile and will serve as a technology base for future multi-dimensional TIP probes. Magnetic field resolution of ±20 gauss and spatial resolution of 5 mm is expected with a system frequency response of 10 MHz. Progress in the development of the orthogonal TIP probe is presented

Additional details

Publishing Information

Publisher
Institute of Electrical and Electronics Engineers, Inc.
Imprint Place
New York, NY (United States)
ISBN
0-7803-2669-5
Imprint Title
IEEE conference record -- abstracts: 1995 IEEE international conference on plasma science
Imprint Pagination
312 p.
Journal Page Range
p. 220-221.

Conference

Title
22. international conference on plasma science.
Dates
5-8 Jun 1995.
Place
Madison, WI (United States).

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
27030395
Subject category
S70: PLASMA PHYSICS AND FUSION TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
FARADAY EFFECT; HOT PLASMA; MAGNETIC FIELDS; MAGNETIC PROBES; MEASURING METHODS; PLASMA DIAGNOSTICS
Descriptors DEC
PLASMA; PROBES

Optional Information

Secondary number(s)
CONF-950612--.