Published October 1, 2017 | Version v1
Journal article

A comparison of the accuracy of the smart sock system to force platform and optical system for measurement of temporal parameters of locomotion

  • 1. Riga Technical University, Institute of Textile and Technology, Riga (Latvia)
  • 2. Riga Technical University, Institute of Bioengineering and Nanotechnology, Riga (Latvia)
  • 3. Latvian Academy of Sport Education, Laboratory of Physical Capacity, Brivibas street 333, Riga (Latvia)

Description

The aim of the study was to present a new DAid®Pressure Sock System for feet locomotion monitoring and to verify it's temporal characteristics by data comparison with the same obtained by two other widely used methods as reference. Designed system is based on sensors which can be knitted directly in the garment or hosiery items. DAid®Pressure Sock System was created for sport and medical applications. Comparison of temporal characteristics of different types of locomotion, obtained using designed system and reference devises, showed good agreement between data. (paper)

Availability note (English)

Available from http://dx.doi.org/10.1088/1757-899X/254/7/072017

Additional details

Publishing Information

Journal Title
IOP Conference Series. Materials Science and Engineering (Online)
Journal Volume
254
Journal Issue
7
Journal Page Range
[6 p.]
ISSN
1757-899X

Conference

Title
17. World Textile Conference - Shaping the Future of Textiles
Acronym
AUTEX 2017
Dates
29-31 May 2017
Place
Corfu (Greece)

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
52061579
Subject category
S36: MATERIALS SCIENCE;
Resource subtype / Literary indicator
Conference
Descriptors DEI
ACCURACY; DESIGN; MONITORING; SENSORS