Published February 1, 1998 | Version v1
Journal article

Development of total uranium analytical method by L X-ray fluorescence

  • 1. Westinghouse Savanah River Co., Aiken, SC (United States). River Technol. Center

Description

This paper describes development of an L X-ray fluorescence technique to perform total uranium analysis using an internal excitation source which is added directly to the sample. The method has been demonstrated with synthetic U samples in the limited concentration range 1-15 g/l, and provides the advantage of simplicity, involving no mechanical parts which would normally be found in an external excitation source. Total uranium is determined by counting L X-rays fluoresced by a microCurie level spike of 109Cd added directly to the sample and without shielding the excitation source from the detector. A method for correction of sample self-absorption is included in the analysis. (orig.)

Additional details

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
403
Journal Issue
2-3
Journal Page Range
p. 383-389.
ISSN
0168-9002
CODEN
NIMAER

Optional Information

Notes
5 refs.