Verification of Mechanical Leaf Gap Error and VMAT Dose Distribution on Varian VitalBeamTM Linear Accelerator
- 1. Seoul National University, Seoul (Korea, Republic of)
- 2. Veterans Health Service Medical Center, Seoul (Korea, Republic of)
Description
The proper position of a multi-leaf collimator (MLC) is essential for the quality of intensity modulated radiation therapy (IMRT) and volumetric modulated arc radiotherapy (VMAT) dose delivery. Task Group (TG) 142 provides a quality assurance (QA) procedure for MLC position. Our study investigated the QA validation of the mechanical leaf gap measurement and the maintenance procedure. Two VitalBeamTM systems were evaluated to validate the acceptance of an MLC position. The dosimetric leaf gaps (DLGs) were measured for 6 MV, 6 MVFFF, 10 MV, and 15 MV photon beams. A solid water phantom was irradiated using 10×10 cm2 field size at source-to-surface distance (SSD) of 90 cm and depth of 10 cm. The portal dose image prediction (PDIP) calculation was implemented on a treatment planning system (TPS) called EclipseTM. A total of 20 VMAT plans were used to confirm the accuracy of dose distribution measured by an electronic portal imaging device (EPID) and those predicted by VMAT plans. The measured leaf gaps were 0.30 mm and 0.35 mm for VitalBeam 1 and 2, respectively. The DLG values decreased by an average of 6.9% and 5.9% after mechanical MLC adjustment. Although the passing rates increased slightly, by 1.5% (relative) and 1.2% (absolute) in arc 1, the average passing rates were still within the good dose delivery level (>95%). Our study shows the existence of a mechanical leaf gap error caused by a degenerated MLC motor. This can be recovered by reinitialization of MLC position on the machine control panel. Consequently, the QA procedure should be performed regularly to protect the MLC system
Additional details
Publishing Information
- Journal Title
- Progress in Medical Physics
- Journal Volume
- 29
- Journal Issue
- 2
- Series
- 16 refs, 3 figs, 4 tabs
- Journal Page Range
- p. 66-72
- ISSN
- 2508-4445
INIS
- Country of Publication
- Korea, Republic of
- Country of Input or Organization
- Korea, Republic of
- INIS RN
- 53092025
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ACCURACY; COLLIMATORS; CONTROL; DOSES; EQUIPMENT; IRRADIATION; LINEAR ACCELERATORS; PHANTOMS; PHOTON BEAMS; QUALITY ASSURANCE; RADIATION DOSE DISTRIBUTIONS; RADIOTHERAPY; SURFACES; VERIFICATION
- Descriptors DEC
- ACCELERATORS; BEAMS; MANAGEMENT; MEDICINE; MOCKUP; NUCLEAR MEDICINE; QUALITY MANAGEMENT; RADIOLOGY; STRUCTURAL MODELS; THERAPY