X-ray film has been used for the dosimetry of intensity modulated radiation therapy (IMRT). However, the over-response of the film to low-energy photons is a significant problem in photon beam dosimetry, especially in regions outside penumbra. In IMRT, the radiation field consists of multiple small fields and their outside-penumbra regions; thus, the film dosimetry, for it involves the source of over-response in its radiation field. In this study we aim to verify and possibly improve film dosimetry for IMRT. Two types of modulated beams were constructed by combining five to seven different static radiation fields using 6 MV x rays. For verifying film dosimetry, x-ray films and an ion chamber were used to measure dose profiles at various depths in a phantom. The film setups include both parallel and perpendicular arrangements against the beam incident direction. In addition, to reduce an over-response, we placed 0.01 in. (0.25 mm) thick lead filters on both sides of the film. Compared with ion-chamber measurement, measured dose profiles showed the film over-response at outside-penumbra and low-dose regions. The error increased with depths and approached 15% as a maximum for the field size of 15 cmx15 cm at 10 cm depth. The use of filters reduced the error down to 3%. In this study we demonstrated that film dosimetry for IMRT involves sources of error due to its over-response to low-energy photons, with the error most transparent in the low-dose region. The use of filters could enhance the accuracy in film dosimetry for IMRT. In this regard, the use of an optimal filter condition is recommended

Additional details

Identifiers

Publishing Information

Journal Title
Medical Physics
Journal Volume
29
Journal Issue
3
Journal Page Range
p. 351-355
ISSN
0094-2405
CODEN
MPHYA6

INIS

Country of Publication
United States
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
35004184
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE; S61: RADIATION PROTECTION AND DOSIMETRY;
Descriptors DEI
DOSIMETRY; FILM DOSIMETRY; OPTIMIZATION; PERSONNEL DOSIMETRY; PHOTON BEAMS; RADIOTHERAPY; X-RAY DETECTION
Descriptors DEC
BEAMS; DETECTION; DOSIMETRY; MEDICINE; NUCLEAR MEDICINE; RADIATION DETECTION; RADIOLOGY; THERAPY

Optional Information

Notes
(c) 2002 American Association of Physicists in Medicine.