Published September 11, 1998 | Version v1
Journal article

Standardization of 139Ce by the liquid scintillation counting using the triple to double coincidence ratio method

  • 1. Radioisotope Centre, Otwock-Swierk, (Poland)
  • 2. Laboratoire Primaire des Rayonnements Ionisants, CEA/BNM, Saclay, BP52, Gif-sur-Yvette (France)

Description

A good agreement of the results of standardization of 139Ce by the TDCR method using RC and LPRI LS-counters was obtained. The results were confirmed by the 4πβ(LS)-γ coincidence, 4πγ(NaI), 4πβ(PC)-γ coincidence and X and γ spectrometer methods. The detection efficiency determination was improved by calculation of the Compton effect scintillation spectrum including the scattered photons events. A substantial modification of the TDCR method was introduced by a new parameter K3=εT/εD3 being the triple to double in third power counting efficiency ratio

Additional details

Identifiers

PII
S0969804397100161;

Publishing Information

Journal Title
Applied Radiation and Isotopes
Journal Volume
49
Journal Issue
9-11
Journal Page Range
p. 1035-1040
ISSN
0969-8043
CODEN
ARISEF

Optional Information

Copyright
Copyright (c) 1998 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.