An investigation of the aseptic loosening of an AISI 316L stainless steel hip prosthesis

  • 1. Institute of Metals and Technology, Lepi pot 11, 1000 Ljubljana (Slovenia)
  • 2. Orthopaedic Clinic Ljubljana, Zaloska 9, 1000 Ljubljana (Slovenia)

Description

The total replacement of joints by the implantation of permanently indwelling prosthetic components has been one of the major successes of modern surgery in terms of relieving pain and correcting deformity. However, the aseptic loosening of a prosthetic-joint component is the most common reason for joint-revision surgery. Furthermore, it is thought that wear particles are one of the major contributors to the development and perpetuation of aseptic loosening. The aim of the present study was to identify the factors related to the aseptic loosening of an AISI 316L stainless steel total hip prosthesis. The stem was evaluated by x-ray photoelectron spectroscopy, with polished and rough regions being analyzed in order to establish the differences in the chemical compositions of both regions. Specific areas were examined using scanning electron microscopy with energy dispersive x-ray spectroscopy and light microscopy.

Availability note (English)

Available from http://dx.doi.org/10.1088/1748-6041/5/4/045012

Additional details

Identifiers

DOI
10.1088/1748-6041/5/4/045012;
PII
S1748-6041(10)56345-9;

Publishing Information

Journal Title
Biomedical Materials (Bristol. Online)
Journal Volume
5
Journal Issue
4
Journal Page Range
[8 p.]
ISSN
1748-605X