Published January 1985
| Version v1
Journal article
Continuous wave intracavity dye laser absorption with an inductively coupled plasma
Description
It is demonstrated that a continuous wave dye laser is capable of operation in an extended cavity fashion while an ICP is running normally intracavity. In the present system, an ICA enhancement factor of 2.5 X 104 for Na is measured as compared to extracavity absorption. For sodium, the detection limit is 0.6 ng/mL using optogalvanic-resonance detection. Due to the limited wavelength coverage of the device, detection limits for five other elements are about 1 μg/mL. The predominant noise source was determined to be wavelength fluctuations in the dye laser output caused by ICP turbulence. The possibility of performing ICP, or other plasma, diagnostics with CW ICA is addressed. 28 references, 2 figures, 1 table
Additional details
Publishing Information
- Journal Title
- Anal. Chem. (Wash.)
- Journal Volume
- 57
- Journal Issue
- 1
- Series
- Anal. Chem. (Wash.).
- Journal Page Range
- 13-16
- ISSN
- 0003-2700
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 16063062
- Subject category
- S37: INORGANIC, ORGANIC, PHYSICAL AND ANALYTICAL CHEMISTRY;
- Resource subtype / Literary indicator
- Numerical Data
- Descriptors DEI
- ABSORPTION SPECTROSCOPY; DYE LASERS; ELECTRO-OPTICAL EFFECTS; EXPERIMENTAL DATA; QUANTITATIVE CHEMICAL ANALYSIS; SODIUM; TRACE AMOUNTS; TURBULENCE
- Descriptors DEC
- ALKALI METALS; AMPLIFIERS; CHEMICAL ANALYSIS; DATA; ELEMENTS; EQUIPMENT; INFORMATION; LASERS; METALS; NUMERICAL DATA; SPECTROSCOPY