Published January 1985 | Version v1
Journal article

Continuous wave intracavity dye laser absorption with an inductively coupled plasma

  • 1. Los Alamos National Lab., NM

Description

It is demonstrated that a continuous wave dye laser is capable of operation in an extended cavity fashion while an ICP is running normally intracavity. In the present system, an ICA enhancement factor of 2.5 X 104 for Na is measured as compared to extracavity absorption. For sodium, the detection limit is 0.6 ng/mL using optogalvanic-resonance detection. Due to the limited wavelength coverage of the device, detection limits for five other elements are about 1 μg/mL. The predominant noise source was determined to be wavelength fluctuations in the dye laser output caused by ICP turbulence. The possibility of performing ICP, or other plasma, diagnostics with CW ICA is addressed. 28 references, 2 figures, 1 table

Additional details

Publishing Information

Journal Title
Anal. Chem. (Wash.)
Journal Volume
57
Journal Issue
1
Series
Anal. Chem. (Wash.).
Journal Page Range
13-16
ISSN
0003-2700

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
16063062
Subject category
S37: INORGANIC, ORGANIC, PHYSICAL AND ANALYTICAL CHEMISTRY;
Resource subtype / Literary indicator
Numerical Data
Descriptors DEI
ABSORPTION SPECTROSCOPY; DYE LASERS; ELECTRO-OPTICAL EFFECTS; EXPERIMENTAL DATA; QUANTITATIVE CHEMICAL ANALYSIS; SODIUM; TRACE AMOUNTS; TURBULENCE
Descriptors DEC
ALKALI METALS; AMPLIFIERS; CHEMICAL ANALYSIS; DATA; ELEMENTS; EQUIPMENT; INFORMATION; LASERS; METALS; NUMERICAL DATA; SPECTROSCOPY