Published September 2008 | Version v1
Journal article

Intercomparison of characterization techniques of filter radiometers in the ultraviolet region

  • 1. Atomic Energy Commission, Damascus (Syrian Arab Republic). Dept. of Protection and Safety

Description

Narrow-band filter radiometers at 248 nm, 313 nm, 330 nm and 368 nm wavelengths were used to compare calibration facilities of spectral (irradiance) responsivity at HUT, NPL and BNM-INM. The results are partly in agreement within the stated uncertainties. Use of demanding artefacts in the intercomparison revealed that the wavelength scales of the participating institutes deviate more than expected. Such effects cannot be seen in typical intercomparisons of spectral responsivity or spectral transmittance, where spectrally neutral samples are used.(author)

Additional details

Publishing Information

Journal Title
Aalam Al-Zarra
Journal Issue
117
Journal Page Range
p. 60
ISSN
1607-985X

INIS

Country of Publication
Syrian Arab Republic
Country of Input or Organization
Syrian Arab Republic
INIS RN
39110134
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
CALIBRATION; FILTERS; RADIATION DETECTORS; RADIOMETERS; ULTRAVIOLET RADIATION; ULTRAVIOLET SPECTRA; WAVELENGTHS
Descriptors DEC
ELECTROMAGNETIC RADIATION; MEASURING INSTRUMENTS; RADIATION DETECTORS; RADIATIONS; SPECTRA

Optional Information

Notes
Abstract of Scientific Research