Quality control and radioprotection in dental cone beam computed tomography - case study
Creators
- 1. Instituto Militar de Engenharia (IME), Rio de Janeiro, RJ (Brazil)
Description
The radiological protection in medical and odontologic radiology follows The Order (Portaria) 453/98 of the Ministry of Health, which presents the minimum set of tests for the constancy X-ray equipment. These tests follow the procedures set forth in the Resolution no. 64, the National Agency for Sanitary Vigilance. This work aims to show a study on dental cone beam computed tomography (CBCT), evaluating the physical parameters that influence the performance and image quality and presenting the appropriate tests to this new system. The authors analyzed the tests specific for computed tomography (CT) of the Resolution no. 64, feasibility assessment of them and if their interpretations are compatible with CBCT. Once determined if testing is feasible, compare with those presented in the manual provided by the equipment manufacturer. The CT scanner used was the Mini-Cat Tomography Scanner Xoran Technologies of KAVO. In the study it was verified that four tests could be reproduced in CBCT: noise, accuracy and uniformity in the number of CT of water and spatial resolution. Considering experimental data, the methodology and tolerance of manufacturer for the first two tests were more appropriate. For the uniformity test of the CT number, we recommend using the phantom quality control. Three new tests were suggested to be made in the quality control of the Cone Beam: linearity, artifacts and alignment of the beam. (author)
Files
43048803.pdf
Files
(435.8 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:4bae5e1c6b81eb69dd61e9a5261ca63a
|
435.8 kB | Preview Download |
Additional details
Publishing Information
- Imprint Pagination
- 10 p.
- Report number
- INIS-BR--10910
Conference
- Title
- International nuclear atlantic conference. Nuclear energy: new jobs for a better life; 17. ENFIR: Meeting on nuclear reactor physics and thermal hydraulics; 10. ENAN: Meeting on nuclear applications; 2. ENIN: Meeting on nuclear industry
- Acronym
- INAC 2011
- Dates
- 24-28 Oct 2011
- Place
- Belo Horizonte, MG (Brazil)
INIS
- Country of Publication
- Brazil
- Country of Input or Organization
- Brazil
- INIS RN
- 43048803
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- BIOLOGICAL MODELS; BIOMEDICAL RADIOGRAPHY; BIOPHYSICS; COMPUTERIZED TOMOGRAPHY; DOSIMETRY; EXTERNAL IRRADIATION; IMAGE PROCESSING; PHANTOMS; RADIATION MONITORING; RADIATION PROTECTION; RADIATION SOURCES; RADIOTHERAPY; TISSUE-EQUIVALENT MATERIALS
- Descriptors DEC
- DIAGNOSTIC TECHNIQUES; IRRADIATION; MATERIALS; MEDICINE; MOCKUP; MONITORING; NUCLEAR MEDICINE; PHYSICS; PROCESSING; RADIOLOGY; STRUCTURAL MODELS; THERAPY; TOMOGRAPHY