Beam feasibility study of a collimator with in-jaw beam position monitors
Creators
- Wollmann, Daniel1
- Nosych, Andriy A.2, 3, 1
- Valentino, Gianluca1
- Aberle, Oliver1
- Aßmann, Ralph W.1
- Bertarelli, Alessandro1
- Boccard, Christian1
- Bruce, Roderik1
- Burkart, Florian1
- Calvo, Eva1
- Cauchi, Marija1
- Dallocchio, Alessandro1
- Deboy, Daniel1
- Gasior, Marek1
- Jones, Rhodri1
- Kain, Verena1
- Lari, Luisella1
- Redaelli, Stefano1
- Rossi, Adriana1
- 1. CERN, Geneva (Switzerland)
- 2. The University of Liverpool, Liverpool (United Kingdom)
- 3. Cockcroft Institute, Daresbury (United Kingdom)
Description
At present, the beam-based alignment of the LHC collimators is performed by touching the beam halo with both jaws of each collimator. This method requires dedicated fills at low intensities that are done infrequently and makes this procedure time consuming. This limits the operational flexibility, in particular in the case of changes of optics and orbit configuration in the experimental regions. The performance of the LHC collimation system relies on the machine reproducibility and regular loss maps to validate the settings of the collimator jaws. To overcome these limitations and to allow a continuous monitoring of the beam position at the collimators, a design with jaw-integrated Beam Position Monitors (BPMs) was proposed and successfully tested with a prototype (mock-up) collimator in the CERN SPS. Extensive beam experiments allowed to determine the achievable accuracy of the jaw alignment for single and multi-turn operation. In this paper, the results of these experiments are discussed. The non-linear response of the BPMs is compared to the predictions from electromagnetic simulations. Finally, the measured alignment accuracy is compared to the one achieved with the present collimators in the LHC
Availability note (English)
Available from http://dx.doi.org/10.1016/j.nima.2014.09.024Additional details
Identifiers
- DOI
- 10.1016/j.nima.2014.09.024;
- PII
- S0168-9002(14)01025-0;
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
- Journal Volume
- 768
- Journal Page Range
- p. 62-68
- ISSN
- 0168-9002
- CODEN
- NIMAER
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 46125411
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ACCURACY; ALIGNMENT; BEAM MONITORS; BEAM POSITION; BEAMS; CERN LHC; CERN SPS SYNCHROTRON; COLLIMATORS; COMPARATIVE EVALUATIONS; FEASIBILITY STUDIES; FLEXIBILITY; NONLINEAR PROBLEMS; OPTICS; SIMULATION
- Descriptors DEC
- ACCELERATORS; CYCLIC ACCELERATORS; EVALUATION; MEASURING INSTRUMENTS; MECHANICAL PROPERTIES; MONITORS; STORAGE RINGS; SYNCHROTRONS; TENSILE PROPERTIES
Optional Information
- Copyright
- Copyright (c) 2014 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.