A software counting loss correction method for neutron activation analysis with short-lived radionuclides
Description
A method has been developed for the correction of counting losses in NAA for the case of a mixture of short-lived radionuclides. It is applicable to systems with Ge detectors and Wilkinson or successive approximation ADC's and will correct losses from pulse pileup and ADC dead time up to 90%. The losses are modeled as a constant plus time-dependent terms expressed as a fourth order polynomial function of the count rates of the short-lived radionuclides. The correction factors are calculated iteratively using the peak areas of the short-lived radionuclides in the spectrum and the average losses as given by the difference between the live time and true time clocks of the MCA. To calibrate the system a measurement is performed for each short-lived nuclide. In a test where the dead time varied from 70% at the start of the measurement to 13% at the end, the measured activities were corrected with an accuracy of 1%. (author) 7 refs.; 2 figs.; 2 tabs
Additional details
Publishing Information
- Journal Title
- Journal of Radioanalytical and Nuclear Chemistry
- Journal Volume
- 193
- Journal Issue
- 2
- Journal Page Range
- p. 239-245.
- ISSN
- 0236-5731
- CODEN
- JRNCDM
Conference
- Title
- 3. International conference on methods and applications of radioanalytical chemistry. Pt.4.
- Acronym
- MARC-III
- Dates
- 10-16 Apr 1994.
- Place
- Kailua-Kona, HI (United States).
INIS
- Country of Publication
- Hungary
- Country of Input or Organization
- Hungary
- INIS RN
- 26072620
- Subject category
- S37: INORGANIC, ORGANIC, PHYSICAL AND ANALYTICAL CHEMISTRY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ACCURACY; CORRECTIONS; DEAD TIME; LIFETIME; LOSSES; MINUTES LIVING RADIOISOTOPES; NEUTRON ACTIVATION ANALYSIS; PULSE PILEUP; RADIATION DETECTORS; SIMULATION
- Descriptors DEC
- ACTIVATION ANALYSIS; CHEMICAL ANALYSIS; ISOTOPES; MEASURING INSTRUMENTS; NONDESTRUCTIVE ANALYSIS; RADIOISOTOPES; TIMING PROPERTIES