Published November 2008 | Version v1
Journal article

Semi-automated measurement of hyperdense, hypodense and heterogeneous hepatic metastasis on standard MDCT slices. Comparison of semi-automated and manual measurement of RECIST and WHO criteria

  • 1. University Hospital, RWTH Aachen University, Department of Diagnostic Radiology, Aachen (Germany)
  • 2. University Hospital, RWTH Aachen University, Institute of Medical Statistics, Aachen (Germany)
  • 3. Siemens Medical Solutions, Forchheim (Germany)
  • 4. University Hospital, RWTH Aachen University, Medical Clinic III, Aachen (Germany)

Description

As semi-automated measurement would be desirable for lesion quantification and therapy-response control, the purpose of this study was to compare semi-automated measurements with manual assessment of different types of hepatic metastases. Seventy-six patients with known liver metastases were analysed. All of them underwent contrast-enhanced 16-MDCT (16 x 0.75 mm collimation, 120 kV, 0.5 s rotation time, 160 mAseff) for evaluation of follow-up status. On the basis of standard reconstructed 5-mm slices (in 4-mm increments), each lesion was quantified based on RECIST and WHO criteria using a semi-automated software tool (Syngo Oncology) and also manually by an experienced radiologist. Results from the software were compared to manual measurements. Statistical analysis was performed applying the concordance correlation coefficient, and results were represented graphically in Bland-Altman plots. A total of 52 hyperdense, 57 hypodense and 56 heterogeneous metastases were found and correctly measured by the software. All three lesion types revealed a strong correlation agreement between measurement techniques [RECIST diameter: 0.93 (hyperdense), 0.95(hypodense), 0.94 (heterogeneous); WHO area: 0.95, 0.98, 0.93]. Semi-automatic measurement of hyperdense, hypodense and heterogeneous liver metastases showed reliable results on standard axial reconstructions in comparison to manual quantification. (orig.)

Availability note (English)

Available from: http://dx.doi.org/10.1007/s00330-008-1050-6

Additional details

Identifiers

Publishing Information

Journal Title
European Radiology
Journal Volume
18
Journal Issue
11
Journal Page Range
p. 2456-2465
ISSN
0938-7994
CODEN
EURAE3

INIS

Country of Publication
Germany
Country of Input or Organization
Germany
INIS RN
39118897
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
COMPUTERIZED TOMOGRAPHY; DIAGNOSIS; LIVER; METASTASES; NEOPLASMS
Descriptors DEC
BODY; DIAGNOSTIC TECHNIQUES; DIGESTIVE SYSTEM; DISEASES; GLANDS; ORGANS; TOMOGRAPHY