Published October 21, 2009 | Version v1
Journal article

Evaluation of electrical crosstalk in high-density photodiode arrays for X-ray imaging applications

  • 1. Detection Technology Inc., FIN-02170 Espoo (Finland)
  • 2. Helsinki University of Technology, FIN-02150 Espoo (Finland)

Description

Electrical crosstalk is one of the important parameters in the photodiode array detector for X-ray imaging applications, and it becomes more important when the density of the photodiode array becomes higher. This paper presents the design of the high-density photodiode array with 250 μm pitch and 50 μm gap. The electrical crosstalk of the demonstrated samples is evaluated and compared with different electrode configurations: cathode bias mode and anode bias mode. The measurement results show good electrical crosstalk, ∼0.23%, in cathode bias mode regardless of the bias voltage, and slightly decreased or increased electrical crosstalk in anode bias mode. Moreover, the quantum efficiency is also evaluated from the same samples, and it behaves similar to the electrical crosstalk. Finally, some design guidance of the high-density photodiode array is given based on the discussion.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.nima.2009.05.060

Additional details

Identifiers

DOI
10.1016/j.nima.2009.05.060;
PII
S0168-9002(09)00995-4;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
610
Journal Issue
1
Journal Page Range
p. 28-30
ISSN
0168-9002
CODEN
NIMAER

Conference

Title
5. international conference on new developments in photodetection
Acronym
NDIP08
Dates
15-20 Jun 2008
Place
Aix-les-Bains (France)

INIS

Country of Publication
Netherlands
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
41063276
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
ANODES; CATHODES; COMPUTERIZED TOMOGRAPHY; CONFIGURATION; DENSITY; DESIGN; ELECTRIC POTENTIAL; EVALUATION; QUANTUM EFFICIENCY; X RADIATION
Descriptors DEC
DIAGNOSTIC TECHNIQUES; EFFICIENCY; ELECTRODES; ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; PHYSICAL PROPERTIES; RADIATIONS; TOMOGRAPHY

Optional Information