Published June 2014 | Version v1
Journal article

Development of adjustable head holder couch in H and H cancer radiation therapy

  • 1. Dept. of Radiation Oncology, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul (Korea, Republic of)

Description

In case of all patients who receive radiation therapy, a treatment plan is established and all steps of treatment are planned in the same geometrical condition. In case of head and neck cancer patients who undergo simulated treatment through computed tomography (CT), patients are fixed onto a table for planning, but laid on the top of the treatment table in the radiation therapy room. This study excogitated and fabricated an adjustable holder for head and neck cancer patients to fix patient's position and geometrical discrepancies when performing radiation therapy on head and neck cancer patients, and compared the error before and after adjusting the position of patients due to difference in weight to evaluate the correlation between patients' weight and range of error. Computed tomography system(High Advantage, GE, USA) is used for phantom to maintain the supine position to acquire the images of the therapy site for IMRT. IMRT 4MV X-rays was used by applying the LINAC(21EX, Varian, U.S.A). Treatment planning system (Pinnacle, ver. 9.1h, Philips, Madison, USA) was used. The setup accuracy was compared with each measurement was repeated five times for each weight (0, 15, and 30Kg) and CBCT was performed 30 times to find the mean and standard deviation of errors before and after the adjustment of each weight. SPSS ver.19.0(SPSS Inc., Chicago, IL,USA) statistics program was used to perform the Wilcoxon Rank test for significance evaluation and the Spearman analysis was used as the tool to analyze the significance evaluation of the correlation of weight. As a result of measuring the error values from CBCT before and after adjusting the position due to the weight difference, X,Y,Z axis was 0.4±0.8 mm, 0.8±0.4 mm, 0 for 0Kg before the adjustment. In 15Kg CBCT before and after adjusting the position due to the weight difference, X,Y,Z axis was 0.2±0.8 mm, 1.2±0.4 mm, 2.0±0.4 mm. After adjusting position was X,Y,Z axis was 0.2±0.4, 0.4±0.5 mm, 0.4±0.5 mm. In 30Kg CBCT before and after adjusting the position due to the weight difference, X,Y,Z axis was 0.8±0.4 mm, 2.4±0.5 mm, 4.4±0.8 mm. After adjusting position was X,Y,Z axis was 0.6±0.5 mm, 1.0±0 mm, 0.6±0.5 mm. When the holder for the head and neck cancer was used to adjust the ab.0ove error value, the error values from CBCT were 0.2±0.8 mm for the X axis, 0.40±0.54 mm for Y axis, and 0 for Z axis. As a result of statistically analyzing each value before and after the adjustment the value was significant with p<0.034 at the Z axis with 15Kg of weight and with p<0.038 and p<0.041 at the Y and Z axes respectively with 30Kg of weight. There was a significant difference with p<0.008 when the analysis was performed through Kruscal-Wallis in terms of the difference in the adjusted values of the three weight groups. As it could reduce the errors, patients' reproduction could be improved for more precise and accurate radiation therapy. Development of an adjustable device for head and neck cancer patients is significant because it improves the reproduction of existing equipment by reducing the errors in patients' position

Additional details

Publishing Information

Journal Title
Journal of the Korean Society for Radiotherapeutic Technology
Journal Volume
26
Journal Issue
1
Series
19 refs, 4 figs, 1 tab
Journal Page Range
p. 43-50
ISSN
1598-8449

INIS

Country of Publication
Korea, Republic of
Country of Input or Organization
Korea, Republic of
INIS RN
47086794
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
COMPUTERIZED TOMOGRAPHY; HEAD; NECK; NEOPLASMS; PATIENTS; PLANNING; RADIOTHERAPY
Descriptors DEC
BODY; DIAGNOSTIC TECHNIQUES; DISEASES; MEDICINE; NUCLEAR MEDICINE; RADIOLOGY; THERAPY; TOMOGRAPHY