OCTAVIUS 4D: Commissioning and routine clinical use of a device for VMAT verifications

Description

Purpose: To commission OCTAVIUS 4D and its arrays 729 and 1000 SRS for VMAT patient-specific QA verifications, and to provide clinical results obtained with this device. Methods: The performance of both arrays was analysed by evaluating: homogeneity, linearity, reproducibility and leakage current. The accuracy of the dose reconstruction of OCTAVIUS 4D with array 729 was evaluated through ion chamber measurements at several points of interest. The performance of both arrays was compared for the same clinical plans. Finally, results of the 3D gamma analysis performed in our hospital for more than 900 patient plans are provided. Results: Both arrays present good homogeneity, linearity, reproducibility, and negligible leakage current. Dose differences between OCTAVIUS-calculated doses and ion chamber measurements or TPS-calculated doses were within 2% for single field plans, and within 4% for clinical plans. The mean value of the local gamma passing rate (3%/3 mm) of clinical plans verified with array 729 is 89.7%, whereas for 1000 SRS it is 97.8%. Conclusions: OCTAVIUS 4D provides accurate 3D dose reconstructions from 2D dose measurements. Clinical results here presented may be used to establish reference levels for VMAT verifications with this device. (Author)

Availability note (English)

Availability www.sefm.es

Additional details

Additional titles

Original title (Spanish)
OCTAVIUS 4D: establecimiento del estado de referencia inicial y uso en rutina clinica de un dispositivo para verifiicaciones VMAT

Publishing Information

Journal Title
Revista de Fisica Medica
Journal Volume
16
Journal Issue
3
Journal Page Range
p. 47-56
ISSN
1576-6632

INIS

Country of Publication
Spain
Country of Input or Organization
Spain
INIS RN
47061472
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
DIAGNOSTIC USES; DOSES; EVALUATION; GAMMA DETECTION; NUCLEAR MEDICINE; PATIENTS; VERIFICATION
Descriptors DEC
DETECTION; MEDICINE; RADIATION DETECTION; USES

Optional Information

Notes
21 refs.