Published February 1, 1986 | Version v1
Journal article

The central electromagnetic calorimeter of UA1

  • 1. CEA Centre d'Etudes Nucleaires de Saclay, 91 - Gif-sur-Yvette (France). Dept. de Physique des Particules Elementaires
  • 2. Oesterreichische Akademie der Wissenschaften, Vienna. Inst. fuer Hochenergiephysik
  • 3. European Organization for Nuclear Research, Geneva (Switzerland)
  • 4. Harvard Univ., Cambridge, MA (USA). Dept. of Physics

Description

We describe the construction, calibration and performance of the central electromagnetic calorimeter of the UA1 experiment at the CERN proton-antiproton collider. The calorimeter is of the lead-scintillator sandwich type. It is 26.4 radiation lengths thick and covers a surface of about 50 m2. We estimate the resolution of the calorimeter for electrons of energy greater than 1 GeV to be the sum in quadrature of 15%/√E (E in GeV) and a constant 3%. The first term comes from the inherent resolution of the calorimeter due to sampling fluctuations and photostatistics. The second term comes from uncertainties in the calibration procedure and dominates the resolution for electrons from W and Z0 decay. The uncertainty in the overall energy scale also reflects the uncertainties in the calibration procedure and is estimated to be 3%. (orig.)

Additional details

Publishing Information

Journal Title
Nucl. Instrum. Methods Phys. Res., Sect. A
Journal Volume
243
Journal Issue
1
Series
Nucl. Instrum. Methods Phys. Res., Sect. A.
Journal Page Range
45-57
ISSN
0168-9002
CODEN
NIMAE

Optional Information

Notes
Also published as report DPhPE--85-07.