Development and evaluation of micro XRF instrument with hybrid X-ray optics

  • 1. Osaka City Univ., Graduate School of Engineering, Dept. of Applied Chemistry and Bioengineering, Osaka (Japan)

Description

We developed a micro-XRF instrument with a hybrid X-ray optic in order to obtain an intense micro X-ray beam in laboratory. The hybrid X-ray optic consists of the polycapillary X-ray lens and the Zr pinhole. The Zr pinhole, which was inserted between the X-ray tube and the polycapillary X-ray lens, was useful for reducing the spot size of micro X-ray beam. In addition, the divergence of micro X-ray beam was reduced by applying the Zr pinhole with a diameter of 400 μm. Moreover, SB ratio of X-ray fluorescence was also improved by using the Zr pinhole. (author)

Additional details

Publishing Information

Journal Title
X-sen Bunseki No Shinpo
Journal Volume
44
Journal Page Range
p. 111-119
ISSN
0911-7806