Thermal and fast neutron dose equivalent distribution measurement of 15-MV linear accelerator using a CR-39 nuclear track detectors
- 1. Medical Physics Research Center, Mashhad University of Medical Sciences, Mashhad (Iran, Islamic Republic of)
- 2. Medical Physics Department, Faculty of Medicine, Gonabad University of Medical Sciences, Gonabad (Iran, Islamic Republic of)
- 3. Medical Physics Department, Faculty of Medicine, Semnan University of Medical Sciences, Semnan (Iran, Islamic Republic of)
- 4. Energy Engineering and Physics Department, Amir Kabir University of Technology, Tehran (Iran, Islamic Republic of)
Description
The main purpose of this study is to measure the contribution of the thermal and fast neutron dose along the central axis of the 15 MV Elekta Precise linac in a tissue equivalent phantom. In order to achieve this purpose, different points were selected in three field sizes of 5 x 5 cm2, 10 x 10 cm2 and 15 x 15 cm2. Fast and thermal neutrons were measured using CR-39 nuclear track detectors with and without thermal neutron converter of 10B, respectively. According to the results, the fast neutron dose equivalent was decreased as the depth increased (field size 5 x 5, 10 x 10 and 15 x 15 cm2 fall from 0.35 to 0.15, 0.5 to 0.3 and 0.5 to 0.3, respectively). Thermal dose equivalent was increased as the depth increased in the tissue equivalent phantom (field size 5 x 5, 10 x 10 and 15 x 15 cm2 rise from 0.1 to 0.4, 0.4 to 0.8 and 0.4 to 0.9, respectively). In conclusion, at depth <3 cm, most existing neutrons are fast and CR-39 films are sensitive to fast neutrons; therefore, they are more appropriate than thermoluminescent dosemeters in measuring neutron dose equivalent. (authors)
Availability note (English)
Available from doi: http://dx.doi.org/10.1093/rpd/ncaa001Additional details
Identifiers
- DOI
- 10.1093/rpd/ncaa001;
Publishing Information
- Journal Title
- Radiation Protection Dosimetry
- Journal Volume
- 188
- Journal Issue
- 4
- Journal Page Range
- p. 503-507
- ISSN
- 0144-8420
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- France
- INIS RN
- 51108269
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S43: PARTICLE ACCELERATORS; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- BORON 10; DOSE EQUIVALENTS; FAST NEUTRONS; LINEAR ACCELERATORS; NEUTRON CONVERTERS; PHANTOMS; THERMAL NEUTRONS; THERMOLUMINESCENT DOSEMETERS
- Descriptors DEC
- ACCELERATORS; BARYONS; BORON ISOTOPES; DOSEMETERS; ELEMENTARY PARTICLES; FERMIONS; HADRONS; ISOTOPES; LIGHT NUCLEI; LUMINESCENT DOSEMETERS; MEASURING INSTRUMENTS; MOCKUP; NEUTRONS; NUCLEI; NUCLEONS; ODD-ODD NUCLEI; STABLE ISOTOPES; STRUCTURAL MODELS
Optional Information
- Notes
- 12 refs.