Dosimetric analysis on the efect of tumor motion in IMRT for liver cancer: comparison of TomoTherapy and VMAT using the Delta4 Hexa-Motion system

  • 1. Chonnam National University Medical School, Gwangju (Korea, Republic of)

Description

The internal target volume (ITV)-based IMRT method can be used to reduce the efect of respiratory tumor motion. In this study, the efect of tumor motion on the dosimetric accuracy during ITV-based IMRT was investigated. ITV-based IMRT plans for a total of ten cases of liver cancer were prepared in TomoTherapy (Tomo) and volumetric modulated arc therapy (VMAT). Delivery quality assurance (DQA) plans were created using Delta4 phantom for the verifcation of dosimetric accuracy of the established Tomo and VMAT plans. DQA measurements were performed in both the static mode and the Hexa-Motion mode by using the Delta4 Hexa-Motion system in Tomo and VMAT. In the static mode, the average gamma passing rate (GPR) using a 3% dose diference/3 mm distance-to-agreement criteria was 99.61% in Tomo and 99.35% in VMAT, which confrmed sufcient dosimetric accuracy in both IMRT methods. For the Hexa-Motion measurement, the average GRP result of 90.93% in Tomo was higher than the result of 82.58% obtained in VMAT. In addition, the diference in the GPR results between the static and the Hexa-motion modes in Tomo was 8.68%, which is approximately 8% less than the 16.77% diference in VMAT. This study verifes that the dosimetric error of ITV-based IMRT due to the tumor motion was larger in VMAT than in Tomo. Also, Tomo can be considered to be more suitable than the conventional LINAC-based VMAT method in terms of the dosimetric accuracy of ITV-based IMRT for the treatment of a moving tumor

Additional details

Publishing Information

Journal Title
Journal of the Korean Physical Society (Online)
Journal Volume
79
Journal Issue
5
Series
18 refs, 8 figs, 2 tabs
Journal Page Range
p. 477-484
ISSN
1976-8524