NEMA NU-04-based performance characteristics of the LabPET-8(TM) small animal PET scanner

  • 1. Division of Nuclear Medicine and Molecular Imaging, Geneva University Hospital, CH-1211 Geneva (Switzerland)

Description

The objective of this study is to characterize the performance of the preclinical avalanche photodiode (APD)-based LabPET-8(TM) subsystem of the fully integrated trimodality PET/SPECT/CT Triumph(TM) scanner using the National Electrical Manufacturers Association (NEMA) NU 04-2008 protocol. The characterized performance parameters include the spatial resolution, sensitivity, scatter fraction, counts rate performance and image-quality characteristics. The PET system is fully digital using APD-based detector modules with highly integrated electronics. The detector assembly consists of phoswich pairs of Lu1.9Y0.1SiO5 (LYSO) and Lu0.4Gd1.6SiO5 (LGSO) crystals with dimensions of 2 x 2 x 14 mm3 having 7.5 cm axial and 10 cm transverse field of view (FOV). The spatial resolution and sensitivity were measured using a small 22Na point source at different positions in the scanner's FOV. The scatter fraction and count rate characteristics were measured using mouse- and rat-sized phantoms fitted with an18F line source. The overall imaging capabilities of the scanner were assessed using the NEMA image-quality phantom and laboratory animal studies. The NEMA-based radial and tangential spatial resolution ranged from 1.7 mm at the center of the FOV to 2.59 mm at a radial offset of 2.5 cm and from 1.85 mm at the center of the FOV to 1.76 mm at a radial offset of 2.5 cm, respectively. Iterative reconstruction improved the spatial resolution to 0.84 mm at the center of the FOV. The total absolute system sensitivity is 12.74% for an energy window of 250-650 keV. For the mouse-sized phantom, the peak noise equivalent count rate (NECR) is 183 kcps at 2.07 MBq cc-1, whereas the peak true count rate is 320 kcps at 2.5 MBq cc-1 with a scatter fraction of 19%. The rat-sized phantom had a scatter fraction of 31%, with a peak NECR of 67 kcps at 0.23 MBq cc-1 and a peak true count rate of 186 kcps at 0.27 MBq cc-1. The average activity concentration and percentage standard deviation were 126.97 kBq ml-1 and 7%, respectively. The performance of the LabPET-8(TM) scanner was characterized based on the NEMA NU 04-2008 standards. The all in all performance demonstrates that the LabPET-8(TM) system is able to produce high-quality and highly contrasted images in a reasonable time, and as such it is well suited for preclinical molecular imaging-based research.

Availability note (English)

Available from http://dx.doi.org/10.1088/0031-9155/56/20/009

Additional details

Identifiers

DOI
10.1088/0031-9155/56/20/009;
PII
S0031-9155(11)91843-0;

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
56
Journal Issue
20
Journal Page Range
p. 6649-6664
ISSN
0031-9155
CODEN
PHMBA7