Published June 29, 2018 | Version v1
Journal article

Ultrasensitive interdigitated capacitance immunosensor using gold nanoparticles

  • 1. Laboratory of Bioanalysis, Institute of Biochemistry and Biophysics, University of Tehran, Tehran (Iran, Islamic Republic of)
  • 2. Department of Nanotechnology, Faculty of Advance Science and Technology, Pharmaceutical Sciences Branch, Islamic Azad University, Tehran (Iran, Islamic Republic of)
  • 3. MEMS and NEMS Lab, Department of Electrical and Computer Engineering, University of Tehran, Tehran (Iran, Islamic Republic of)
  • 4. Pishtazteb Medical Co., Tehran (Iran, Islamic Republic of)

Description

Immunosensors based on interdigitated electrodes (IDEs), have recently demonstrated significant improvements in the sensitivity of capacitance detection. Herein, a novel type of highly sensitive, compact and portable immunosensor based on a gold interdigital capacitor has been designed and developed for the rapid detection of hepatitis B surface antigen (HBsAg). To improve the efficiency of antibody immobilization and time-saving, a self-assembled monolayer (SAM) of 2-mercaptoethylamine film was coated on IDEs. Afterwards, carboxyl groups on primary antibodies were activated through 1-ethyl-3-(3-dimethylaminopropyl)carbodiimide and were immobilized on amino-terminated SAM for better control of the oriented immobilization of antibodies on gold IDEs. In addition, gold nanoparticles conjugated with a secondary antibody were used to enhance the sensitivity. Under optimal conditions, the immunosensor exhibited the sensitivity of 0.22 nF.pg ml–1, the linear range from 5 pg ml−1 to 1 ng ml–1 and the detection limit of 1.34 pg ml−1, at a signal-to-noise ratio of 3. (paper)

Availability note (English)

Available from http://dx.doi.org/10.1088/1361-6528/aabca3

Additional details

Identifiers

Publishing Information

Journal Title
Nanotechnology (Print)
Journal Volume
29
Journal Issue
26
Journal Page Range
[11 p.]
ISSN
0957-4484