Published February 1973
| Version v1
Journal article
Operational characteristics of a high purity germanium photon spectrometer cooled by a closed-cycle cryogenic refrigerator
Creators
Description
Experience with several (commercially available) miniature closed-cycle refrigerators used to cool a high purity germanium photon detector is reported. While maintaining good resolution, closed-cycle cooling has resulted in a versatile photon spectrometer with long storage times, variable length run times, and freedom from the need for a liquid nitrogen supply. The demonstrated resolution and background levels indicate the concept to be especially useful for field measurements in problems of general radioisotope detection and in some applications of on-line x-ray fluorescence.
Additional details
Identifiers
Publishing Information
- Journal Title
- IEEE Transactions on Nuclear Science
- Journal Volume
- 20
- Journal Issue
- 1
- Series
- IEEE (Inst. Elec. Electron. Eng.), Trans. Nucl. Sci.
- Journal Page Range
- 522-527
- ISSN
- 0018-9499
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 4081170
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- GAMMA SPECTROMETERS; GE SEMICONDUCTOR DETECTORS; PERFORMANCE; REFRIGERATION; X-RAY SPECTROMETERS
- Descriptors DEC
- COOLING; MEASURING INSTRUMENTS; RADIATION DETECTORS; SEMICONDUCTOR DETECTORS; SPECTROMETERS