Published February 1973 | Version v1
Journal article

Operational characteristics of a high purity germanium photon spectrometer cooled by a closed-cycle cryogenic refrigerator

Description

Experience with several (commercially available) miniature closed-cycle refrigerators used to cool a high purity germanium photon detector is reported. While maintaining good resolution, closed-cycle cooling has resulted in a versatile photon spectrometer with long storage times, variable length run times, and freedom from the need for a liquid nitrogen supply. The demonstrated resolution and background levels indicate the concept to be especially useful for field measurements in problems of general radioisotope detection and in some applications of on-line x-ray fluorescence.

Additional details

Identifiers

Publishing Information

Journal Title
IEEE Transactions on Nuclear Science
Journal Volume
20
Journal Issue
1
Series
IEEE (Inst. Elec. Electron. Eng.), Trans. Nucl. Sci.
Journal Page Range
522-527
ISSN
0018-9499

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
4081170
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
GAMMA SPECTROMETERS; GE SEMICONDUCTOR DETECTORS; PERFORMANCE; REFRIGERATION; X-RAY SPECTROMETERS
Descriptors DEC
COOLING; MEASURING INSTRUMENTS; RADIATION DETECTORS; SEMICONDUCTOR DETECTORS; SPECTROMETERS

Optional Information

Notes
Updated automatically by Metadata and Full-Text Enrichment Agent