Evaluation of low contrast resolution using contrast-to-noise ratio in CT
- 1. Nara Medical Univ., Kashihara (Japan). Hospital
Description
We evaluated low contrast resolution in CT images from various apparatuses by using the second recommendation of the x-ray CT commission. This method is based on visual assessment by an observer using a CT performance evaluation phantom. We evaluated CT images with low contrast resolution using the contrast-to-noise ratio (CNR) in terms of quantity of data, region of interest (ROI) size, and the relation between the second recommendation and CNR, and compared various kinds of machines according to CNR. We found that this method had certain factors that influenced the precision of measurement. However, there was a correlation between CNR and visual evaluation, and we were able to evaluate low contrast resolution in each apparatus. This method was considered a useful step, because low contrast resolution could be easily evaluated. (author)
Additional details
Publishing Information
- Journal Title
- Nippon Hoshasen Gijutsu Gakkai Zasshi
- Journal Volume
- 55
- Journal Issue
- 8
- Journal Page Range
- p. 778-782
- ISSN
- 0369-4305
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 30056859
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- CAT SCANNING; COMPARATIVE EVALUATIONS; IMAGES; PERFORMANCE; PHANTOMS; SIGNAL-TO-NOISE RATIO; SPATIAL RESOLUTION
- Descriptors DEC
- COMPUTERIZED TOMOGRAPHY; EVALUATION; MOCKUP; RESOLUTION; STRUCTURAL MODELS; TOMOGRAPHY