Performance testing of Eberline Alpha 6 and Alpha 6A continuous air monitors

Description

Eberline Alpha 6 and Alpha 6A continuous air monitors (CAMs) were tested against the performance criteria of the International Electrotechnical Commission standard 761-6, ''Equipment for Continuously Monitoring Radioactivity in Gaseous Effluents, Part 6: Specific Requirements for Transuranic Aerosol Effluent Monitors'', and against ANSI N42.17B, ''Performance Specification Health Physics Instrumentation--Occupational Airborne Radioactivity Monitoring Instrumentation''. The performance criteria require the CAM's response to a radioactive source to remain within a tolerance while the CAM is exposed to an external influence such as temperature, electromagnetic fields, or ionizing radiations. The CAMs complied within specified tolerances with a majority of the performance specifications. The most significant problems with CAM performance were noted during exposures to external nonionizing radiation fields (radio frequency fields). At numerous frequencies, the CAMs did not respond to radioactive material in the filter holder. At other frequencies and in some orientations, the CAMs overresponded by orders of magnitude. In addition to sensitivity to external nonionizing radiation fields, the CAMs exhibited sensitivity to electrostatic discharges

Availability note (English)

Available from INIS in electronic form; ALSO AVAILABLE FROM OSTI AS DE98053052; NTIS; US GOVT. PRINTING OFFICE DEP.

Files

30009459.pdf

Files (334.1 kB)

Name Size Download all
md5:2cc701bbb4e4c71fc5d347ee70b5da37
334.1 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
22 p.
Report number
PNNL--11527