Published June 1996 | Version v1
Journal article

A 48x48 CdZnTe array with multiplexer readout

  • 1. Univ. of Arizona, Tucson, AZ (United States)

Description

The authors report results of gamma-ray imaging and energy-resolution tests of a 48x48 CdZnTe array. The detectors have 125 microm square pixel electrodes produced by photolithography and are indium-bump-bonded to a multiplexer readout circuit. Using a collimated beam of 140 keV gamma rays of 120 microm diameter centered on one pixel, they found that the majority of events produced significant charge deposition in nearby pixels. Charge and energy are transported out of the pixel by charge diffusion, photoelectron range, Compton scattering, and escape of K x-rays. These effects also distort single-pixel spectra, although photopeaks are still discernible at 140 keV. When signals from neighboring pixels are summed together to correct for this charge spreading, an energy resolution of 10 keV is obtained at 140 keV. Corrections will be simpler and energy resolution should be better for the 380 microm pixels of the 64x64 CdZnTe arrays the authors are constructing for an ultra-high-resolution brain imager

Additional details

Publishing Information

Journal Title
IEEE Transactions on Nuclear Science
Journal Volume
43
Journal Issue
3Pt2
Journal Page Range
p. 1253-1259.
ISSN
0018-9499
CODEN
IETNAE

Conference

Title
IEEE nuclear science symposium and medical imaging conference.
Dates
23-28 Oct 1995.
Place
San Francisco, CA (United States).

Optional Information

Secondary number(s)
CONF-951073--.