Published February 1974
| Version v1
Journal article
Chronic in vivo testing of a fully implantable intracranial pressure sensor
Description
This paper describes the authors' approach to the problem of long term intracranial pressure measurement. Several prior devices have not proven useful for clinical studies of more than a few days duration. Study of 22 of the authors' devices in primates has established the validity of both engineering and medical design assumptions. Useful pressure data has been collected for up to 2 months. Redesign to decrease tambour permeability should allow a useful life of months or years.
Additional details
Identifiers
Publishing Information
- Journal Title
- IEEE Transactions on Nuclear Science
- Journal Volume
- 21
- Journal Issue
- 1
- Series
- IEEE (Inst. Elec. Electron. Eng.), Trans. Nucl. Sci.
- Journal Page Range
- 702-706
- ISSN
- 0018-9499
Conference
- Title
- 20. nuclear science symposium and 5. nuclear power systems symposium.
- Dates
- 14 Nov 1974.
- Place
- San Francisco, CA.
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 5143924
- Subject category
- S07: ISOTOPES AND RADIATION SOURCES;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- HEAD; IN VIVO; PERFORMANCE TESTING; PRESSURE GAGES; PRESSURE MEASUREMENT; RADIOMETRIC GAGES
- Descriptors DEC
- BODY; BODY AREAS; MEASURING INSTRUMENTS; TESTING
Optional Information
- Notes
- Updated automatically by Metadata and Full-Text Enrichment Agent