Published February 1974 | Version v1
Journal article

Chronic in vivo testing of a fully implantable intracranial pressure sensor

  • 1. Medical Coll. of Wisconsin, Milwaukee

Description

This paper describes the authors' approach to the problem of long term intracranial pressure measurement. Several prior devices have not proven useful for clinical studies of more than a few days duration. Study of 22 of the authors' devices in primates has established the validity of both engineering and medical design assumptions. Useful pressure data has been collected for up to 2 months. Redesign to decrease tambour permeability should allow a useful life of months or years.

Additional details

Identifiers

Publishing Information

Journal Title
IEEE Transactions on Nuclear Science
Journal Volume
21
Journal Issue
1
Series
IEEE (Inst. Elec. Electron. Eng.), Trans. Nucl. Sci.
Journal Page Range
702-706
ISSN
0018-9499

Conference

Title
20. nuclear science symposium and 5. nuclear power systems symposium.
Dates
14 Nov 1974.
Place
San Francisco, CA.

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
5143924
Subject category
S07: ISOTOPES AND RADIATION SOURCES;
Resource subtype / Literary indicator
Conference
Descriptors DEI
HEAD; IN VIVO; PERFORMANCE TESTING; PRESSURE GAGES; PRESSURE MEASUREMENT; RADIOMETRIC GAGES
Descriptors DEC
BODY; BODY AREAS; MEASURING INSTRUMENTS; TESTING

Optional Information

Notes
Updated automatically by Metadata and Full-Text Enrichment Agent