Problems encountered when integrating peaks obtained in gas chromatography. Application to the determination of hydrogen isotopes
Description
The gas chromatography of hydrogen isotopes, in use for twelve years, is easy to adapt to gloves boxes. The long holding times lead to an analysis duration of around 30 minutes. This is why we have low resolution mass spectrometers. These appliances quickly respond to analytical requirements and enable an isotope processing installation to be controlled in real time. On the occasion of the use of this new analytical method, the effect of the integrators employed in chromatography on the accuracy of the results was noticed. The integrator does not solve all chromatography problems; the chromatograph-integrator must be considered as a whole. In particular what kind of integrator should be chosen in the presence of badly separated and dissymmetrical peaks. This is the question that must be posed and for which this study provides an answer
Availability note (English)
MF available from INIS under the Report Number.Abstract (French)
Mise en oeuvre depuis une douzaine d'annees, la chromatographie en phase gazeuse des isotopes de l'hydrogene est facile a adapter en boite a gants. Les temps de retention longs conduisent a une duree d'analyse de 30 minutes environ. C'est la raison pour laquelle nous nous sommes equipes de spectrometres de masse basse resolution. Ces appareils repondent rapidement aux demandes d'analyses et permettent le pilotage en temps reel d'une installation de traitement des isotopes. A l'occasion de la mise en oeuvre de cette nouvelle methode d'analyse, nous avons remarque l'influence des integrateurs utilises en chromatographie sur la justesse des resultats. L'integrateur ne resout pas tous les problemes de chromatographie; il faut considerer l'ensemble chromatographe-integrateur. En particulier quel type d'integration choisir en presence de pics mal separes et dissymetriques. C'est la question qu'il convient de se poser et a laquelle nous repondons dans cette etudeFiles
14774738.pdf
Files
(202.3 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:2491fcb7175397f6144caf721ff9095b
|
202.3 kB | Preview Download |
Additional details
Additional titles
- Original title (French)
- Problemes rencontres lors de l'integration des pics obtenus en chromatographie en phase gazeuse. Application au dosage des isotopes de l'hydrogene
Publishing Information
- Imprint Pagination
- 13 p.
- Report number
- CEA-CONF--6559
Conference
- Title
- CETAMA meeting on gas phase chromatography.
- Dates
- 21-22 Oct 1982.
- Place
- Fontenay-aux-Roses (France).
INIS
- Country of Publication
- France
- Country of Input or Organization
- France
- INIS RN
- 14774738
- Subject category
- S37: INORGANIC, ORGANIC, PHYSICAL AND ANALYTICAL CHEMISTRY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ERRORS; GAS CHROMATOGRAPHY; HYDROGEN ISOTOPES; ISOTOPE RATIO; MASS RESOLUTION; PEAKS
- Descriptors DEC
- CHROMATOGRAPHY; RESOLUTION; SEPARATION PROCESSES