Published November 1975 | Version v1
Journal article

An approach to microdosimetry in a π- meson beam using nuclear emulsions

  • 1. Saint Bartholomew's Hospital, London (UK)

Description

K5 emulsions 10 μm thick were exposed at various depths in a perspex phantom to a 70 MeV π- meson beam and counts taken of the tracks in emulsion volumes 7 x 7 x 10 μm3. Data are presented on the number of track events and also the total number of grains associated with each event in each of four categories spanning the LET range of the secondary particles. The number of heavy tracks (category 4) showed an increased incidence in the region of the stopping π- mesons (14.5 cm in perspex) while the number of single grains (category 1) decreased with depth. Categories 2 and 3 (grain clusters and light tracks) were approximately constant with depth. An estimate of the grain sensitivity was obtained by taking the proton as representative of the whole range of secondary particles. This procedure gave a value of 5 keV per grain in the π- peak. The LET of light tracks was therefore in the range 1 to 10 keV μm -1 in emulsion, scaling to 0.4 to 4 keV μm-1 in water. Heavy tracks had LET values in excess of 4 keV μm-1 in water. (author)

Additional details

Identifiers

Publishing Information

Journal Title
Physics in Medicine & Biology
Journal Volume
20
Journal Issue
6
Series
Phys. Med. Biol.
Journal Page Range
918-925
ISSN
0031-9155

Optional Information

Notes
Updated automatically by Metadata and Full-Text Enrichment Agent