Published August 1990 | Version v1
Journal article

Field comparison of several commercially available randon detectors

  • 1. Univ. of Iowa, Institute of Agricultural Medicine and Occupational Health, Iowa City (USA)

Description

To determine the accuracy and precision of commercially available radon detectors in a field setting, 15 detectors from six companies were exposed to randon and compared to a reference randon level. The detectors from companies that had already passed National Radon Measurement Proficiency Program testing had better precision and accuracy than those detectors awaiting proficiency testing. Charcoal adsorption detectors and diffusion barrier charcoal adsorption detectors performed very well, and the latter detectors displayed excellent time averaging ability. Alternatively, charcoal liquid scintillation detectors exhibited acceptable accuracy but poor precision, and bare alpha registration detectors showed both poor accuracy and precision. The mean randon level reported by the bare alpha registration detectors was 68 percent lower than the radon reference level

Additional details

Publishing Information

Journal Title
American Journal of Public Health
Journal Volume
80
Journal Issue
8
Series
Am. J. Public Health.
Journal Page Range
926-930
ISSN
0090-0036
CODEN
AJHEA