Field comparison of several commercially available randon detectors
Creators
- 1. Univ. of Iowa, Institute of Agricultural Medicine and Occupational Health, Iowa City (USA)
Description
To determine the accuracy and precision of commercially available radon detectors in a field setting, 15 detectors from six companies were exposed to randon and compared to a reference randon level. The detectors from companies that had already passed National Radon Measurement Proficiency Program testing had better precision and accuracy than those detectors awaiting proficiency testing. Charcoal adsorption detectors and diffusion barrier charcoal adsorption detectors performed very well, and the latter detectors displayed excellent time averaging ability. Alternatively, charcoal liquid scintillation detectors exhibited acceptable accuracy but poor precision, and bare alpha registration detectors showed both poor accuracy and precision. The mean randon level reported by the bare alpha registration detectors was 68 percent lower than the radon reference level
Additional details
Publishing Information
- Journal Title
- American Journal of Public Health
- Journal Volume
- 80
- Journal Issue
- 8
- Series
- Am. J. Public Health.
- Journal Page Range
- 926-930
- ISSN
- 0090-0036
- CODEN
- AJHEA
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 21086365
- Subject category
- S54: ENVIRONMENTAL SCIENCES; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ACCURACY; AIR POLLUTION; COMPARATIVE EVALUATIONS; RADIATION DETECTORS; RADIATION MONITORING; RADIOECOLOGICAL CONCENTRATION; RADON
- Descriptors DEC
- ECOLOGICAL CONCENTRATION; ELEMENTS; EVALUATION; MEASURING INSTRUMENTS; MONITORING; NONMETALS; POLLUTION; RARE GASES