The use of trans-stilbene scintillators for the measurement of dose in γ-n mixed fields

  • 1. Comitato Nazionale per l'Energia Nucleare, Frascati (Italy). Laboratori Nazionali di Frascati

Description

The possibility of employing a trans-stilbene crystal scintillator in the dosimetry of γ-n mixed fields has been studied. By a suitable linear combination of the slow and fast components of the light pulse, a response can be obtained to protons and to γ-rays, that makes feasible the use of the scintillator for the measurement of the equivalent dose in γ-n mixed field. The energy range that can be covered is from 0.1 up to 5+6 MeV for γ-rays and from 0.1 up to 12+14 MeV for neutrons. It must be noticed that, in mixed radiation field where the contribution of the neutrons to equivalent dose is not much greater than that one of γ-rays, the slow component of light pulse can be assumed as a measure of the absorbed dose. Then, under this conditions, the measurement of the quality factor of the field can be obtained with sufficiently high accuracy

Additional details

Publishing Information

Imprint Title
Proceedings, second symposium on neutron dosimetry in biology and medicine, Neuherberg/Muenchen, September 30-October 4, 1974
Imprint Pagination
p. 613-625.
Report number
EUR--5273(vol.1)

Conference

Title
2. Symposium on neutron dosimetry in biology and medicine.
Dates
30 Sep 1974.
Place
Neuherberg/Muenchen, F.R. Germany.