Published April 18, 2013 | Version v1
Journal article

A Low-power and Compact-sized Wearable Bio-impedance Monitor with Wireless Connectivity

  • 1. Holst Centre / imec, High Tech Campus 31, 5656 AE, Eindhoven (Netherlands)
  • 2. imec, Kapeldreef 75, B-3001, Leuven (Belgium)
  • 3. Philips Research, Eindhoven (Netherlands)

Description

In this paper, we present a new bio-impedance monitor for wearable and continuous monitoring applications. The system consumes less than 14.4mW when measuring impedance, and 0.9mW when idling. Its compact size (4.8cm × 3cm × 2cm) makes it suitable for portable and wearable use. The proposed system has an accuracy of 0.5Ω and resolution of 0.2Ω on both resistance (R) and reactance (X) measurements, for impedance ranging between (j0.7)Ω to (54+j5)Ω with 2.9<φ<5.7. We also report the results of the system validation using passive loads as human tissue model, and show our wireless and miniaturized bio-impedance monitoring system has comparable performances with a reference system.

Availability note (English)

Available from http://dx.doi.org/10.1088/1742-6596/434/1/012013

Additional details

Publishing Information

Journal Title
Journal of Physics. Conference Series (Online)
Journal Volume
434
Journal Issue
1
Journal Page Range
[4 p.]
ISSN
1742-6596

Conference

Title
15. international conference on electrical bio-impedance (ICEBI); 14. conference on electrical impedance tomography (EIT)
Dates
22-25 Apr 2013
Place
Heilbad Heiligenstadt (Germany)

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
44118977
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
COMPACTS; IMPEDANCE; MONITORING; MONITORS; PERFORMANCE
Descriptors DEC
MEASURING INSTRUMENTS