Published 1996
| Version v1
Miscellaneous
A Si(Li) spectrometer with thermoelectric cooling
Creators
- 1. Academia Sinica, Shanghai, SH (China). Shanghai Inst. of Nuclear Research
Description
A Si(Li) spectrometer with thermoelectric cooler has been manufactured. The spectrometer has a high-valued resistor feedback preamplifier. At -50 degree C and a bias of -300 V, the full width at half maximum [FWHM] for 28 mm2 Si(Li) detector is 700 eV approximately for 5.9 keV X ray. The temperature of the detector assembly, detector leakage currents at several bias voltages and resolution have been studied for cooler input power levels over the range of 2.8 to 45 W
Additional details
Publishing Information
- Imprint Title
- Proceedings of 8. national conference on nuclear electronic and nuclear detection technology: Pt.1
- Imprint Pagination
- 348 p.
- Journal Page Range
- p. 23, 25-26.
Conference
- Title
- 8. national conference on nuclear electronics and nuclear detection technology.
- Dates
- 2-7 Dec 1996.
- Place
- Zhuhai, GD (China).
INIS
- Country of Publication
- China
- Country of Input or Organization
- China
- INIS RN
- 28055415
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference, Non-conventional Literature
- Descriptors DEI
- COOLING; ELECTRICAL PROPERTIES; ENERGY RESOLUTION; LI-DRIFTED SI DETECTORS; TEMPERATURE DEPENDENCE; THERMOELECTRIC REFRIGERATORS; TIME DEPENDENCE; X-RAY SPECTROMETERS
- Descriptors DEC
- DIRECT ENERGY CONVERTERS; LI-DRIFTED DETECTORS; MEASURING INSTRUMENTS; PHYSICAL PROPERTIES; RADIATION DETECTORS; REFRIGERATORS; RESOLUTION; SEMICONDUCTOR DETECTORS; SI SEMICONDUCTOR DETECTORS; SPECTROMETERS