Usefulness of DICOM headers in the analysis of two biplane X-ray systems setting (image intensifier and flat panel) used in pediatric interventional cardiology in Chile
- 1. Universidad de Tarapaca, Arica (Chile). Dept. de Ciencias Clinicas
- 2. Universidad Complutense, Madrid, (Spain). Hospital Clinico San Carlos. Dept. de Radiologia
- 3. Hospital Luis Calvo Mackenna, Santiago (Chile). Dept. de Hemodinamia
- 4. Universidad Catolica de Chile, Santiago (Chile). Hospital Clinico. Servicio de Hemodinamia
- 5. Instituto de Salud Publica de Chile, Santiago (Chile)
Description
The setting of two biplane X ray systems were evaluated (image intensifier (II) and flat panel (PP)), through DICOM tags from 32 images created during the characterization of both systems. The technical parameters adjusted for systems were: 63,8 to 80,0 kV and 15,0 to 388,0 mA, for the system with II and 52,0 to 77,0 kV and 25,0 to 476,0 mA, for the system with PP detector. Both equipment presented a different mA adjustment, when moving from fluoroscopy modes low dose (FL), medium dose (FM) and high dose (FH) to cine mode (CI). Two dosimetric quantities were evaluated, the first one was the dose-area product (DAP) which gave as a result for FB mode, between 0,03 to 0,35 uGycm2/image (II) and from 0,05 a 0,69 uGycm2/image (PP), when the polymethyl methacrylate (PMMA) thickness was incremented from 4 to 16 cm. In cine mode the DAP quantity showed, percentage values from 24 to -1 % for the same PMMA increment. Skin cumulative dose was the second quantity evaluated and showed an increment of incident air kerma (KAI)/image in factors from 17 to 35 (II) and 15 to 28 (PP) when used in CI mode instead of FB mode, to the different PMMA thicknesses used. This dose increment for CI mode must be considered by cardiologists, to use the fluoroscopic run as an alternative to document part of the procedures when there is no need to use a high quality image (author)
Additional details
Additional titles
- Original title (Spanish)
- Utilidad de las cabeceras DICOM en el analisis de la configuracion de dos sistemas de rayos X biplano (intensificador de imagen y panel plano) utilizados en cardiologia intervencionista pediatrica en Chile
Publishing Information
- Publisher
- ABFM
- Imprint Place
- Sao Paulo (Brazil)
- ISBN
- 978-85-85205-94-2
- Imprint Title
- Proceedings of the 14. Brazilian congress on medical physics; 4. Symposium on instrumentation and medical imaging
- Imprint Pagination
- [vp.]
- Journal Page Range
- [4 p.]
Conference
- Title
- 14. Brazilian congress on medical physics; 4. Symposium on instrumentation and medical imaging
- Original Conference Title
- 14. Congresso brasileiro de fisica medica
- Dates
- 8-12 Jul 2009
- Place
- Sao Paulo, SP (Brazil)
INIS
- Country of Publication
- Brazil
- Country of Input or Organization
- Brazil
- INIS RN
- 41109812
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- BIOMEDICAL RADIOGRAPHY; BIOPHYSICS; DOSE COMMITMENTS; DOSE EQUIVALENTS; DOSIMETRY; IMAGE CONVERTERS; IMAGE INTENSIFIERS; IMAGE PROCESSING; IRRADIATION; RADIATION DOSE UNITS; RADIATION DOSES; RADIATION PROTECTION; RADIATIONS; X RADIATION; X-RAY EQUIPMENT; X-RAY RADIOGRAPHY
- Descriptors DEC
- DIAGNOSTIC TECHNIQUES; DOSES; ELECTROMAGNETIC RADIATION; EQUIPMENT; IMAGE TUBES; INDUSTRIAL RADIOGRAPHY; IONIZING RADIATIONS; MATERIALS TESTING; MEDICINE; NONDESTRUCTIVE TESTING; NUCLEAR MEDICINE; PHYSICS; PROCESSING; RADIATIONS; RADIOLOGY; TESTING; UNITS
Optional Information
- Notes
- Imprint:Anais do 14. Congresso brasileiro de fisica medica; 4. Simposio de instrumentacao e imagens medicas