Published December 1, 2005 | Version v1
Journal article

1.3kg bolometers to search for rare events

  • 1. Dipartimento di Fisica dell'Universita di Milano-Bicocca and INFN, Sezione di Milano, I-20126 Milan (Italy)
  • 2. INFN, Laboratori Nazionali del Gran Sasso, I-67010 Assergi (L'Aquila) (Italy)
  • 3. Laboratorio de Fisica Nuclear y Altas Energias, Universidad de Zaragoza, 50009 Zaragoza (Spain)

Description

Two TeO2 crystal bolometers of 1.3kg each, the largest single crystals ever operated with this technique, have been successfully realized and tested below 10mK, in a dilution refrigerator located deep underground in the Gran Sasso National Laboratories. The calibration spectrum, obtained using an external 232Th γ-ray source, shows an energy resolution of 3-4keV FWHM from 0.5 to 2.6MeV, for both detectors. In the α region, a 4.3keV FWHM resolution has been observed on the 5407keV peak due to the α decay of 210Po, a natural contaminant of TeO2 crystals

Additional details

Identifiers

DOI
10.1016/j.nima.2005.07.060;
PII
S0168-9002(05)01532-9;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
554
Journal Issue
1-3
Journal Page Range
p. 300-305
ISSN
0168-9002
CODEN
NIMAER

Optional Information

Copyright