Published August 1, 1999 | Version v1
Journal article

The reference peak areas of the 1995 IAEA test spectra for gamma-ray spectrum analysis programs are absolute and traceable

Creators

Description

A previously validated algorithm for absolute peak area determination was used to verify the reference peak areas supplied with the 1995 IAEA test spectra for gamma-ray spectrometry. These reference peak areas turn out to be absolute and traceable to a precision of 0.9%: The reference peak areas are possibly too low by a factor 0.992±0.009. It is proposed to employ the test spectra and reference areas to validate the peak areas obtained with any algorithm in gamma-ray spectrometry. (author)

Additional details

Identifiers

PII
S0168900299002570;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
432
Journal Issue
1
Journal Page Range
p. 74-76