Published October 1980 | Version v1
Journal article

Trial making of a positive drawing phantom and its application to whole-body imaging devices

  • 1. Chiba Univ. (Japan). School of Medicine

Description

In whole-body RI imaging, there are more instances of the positive pictures detecting the radioisotopes accumulating in morbid positions, such as Tc-99m bone scanning. The phantoms used to mutually compare RI imaging devices and to test their performance employ negative drawing targets embedded rather than positive ones. A simple positive drawing phantom has been made for trial, and applying this to whole-body scanning devices, the performance and the target drawing ability under different scanning conditions were comparatively examined. Though similar to Rollo's phantom, the phantom made for positive drawing uses acryl plate for its outer structure and target portions. The positive targets are cylindrical, and the diameters are 2, 4, 6, 8, 10, and 20 mm, and the subject contrasts are 5, 2, 1, 0.5 and 0.2. The aqueous solution of Tc-99m of about 2 mCi was injected into the phantom, and this was scanned with a whole-body camera and a multi-detector type whole-body scanner. With the phantom pictures close to actual clinical condition, the positive drawing phantom is conveniently capable of comparing the respective imaging devices for intended purposes. (J.P.N.)

Additional details

Publishing Information

Journal Title
Eizo Joho, Medikaru
Journal Volume
12
Journal Issue
16
Series
Eizo Joho, Medikaru.
Journal Page Range
1019-1022
ISSN
0389-214X