Published November 2002 | Version v1
Book

The development of flow injection technique for rapid uranium determination in urine samples

  • 1. Japan Atomic Energy Research Inst., Tokai, Ibaraki (Japan). Tokai Research Establishment

Description

Much attention has been paid to precise and rapid determination of actinide elements in various biological samples. The importance of actinide bioassay is caused by their radiological hazard due to the internal exposure. The present research is devoted to the development of a flow injection extraction chromatography (FI-EC) as an analytical procedure for rapid determination of uranium in urine by ICP-MS. The developed FI-EC-ICPMS system includes automated radiochemical separation based on the use of UTEVA·Spec extraction chromatography resin followed by immediate ICP-MS measurement. Microwave digestion is applied to decompose urine samples prior to the FI-EC-ICPMS measurement. The system automatically removes matrix substances, thus, overcomes problem of matrix effect during ICP-MS measurement. Only few cm3 of urine are required for a single uranium analysis because of a low detection limit of the ICP-MS method. The developed FI-EC-ICPMS system is fully automated and virtually does not require handwork. By the FI-EC-ICPMS method uranium determination could be performed within 8 min per urine sample. (author)

Availability note (English)

Available from the Internet at URL https://doi.org/10.1080/00223131.2002.10875529
Part of:
Proceedings of actinide 2001 international conference

Additional details

Publishing Information

Publisher
Atomic Energy Society of Japan
Imprint Place
Tokyo (Japan)
Imprint Title
Proceedings of actinide 2001 international conference
Imprint Pagination
986 p.
Journal Page Range
p. 556-559

Conference

Title
Actinide 2001 international conference
Dates
4-9 Nov 2001
Place
Hayama, Kanagawa (Japan)

Optional Information

Notes
5 refs., 4 figs., 2 tabs.; This record replaces 34044881